Enquire Now
CALsys 15 110 DBC

CALsys -15/110 Peltier Based Temperature Calibrator

Part Code: CALsys -15/110

Temperature Range: -15 to 110°C

Stability: ±0.03°C at -15°C, ±0.05°C at 0°C, ±0.07°C at 110°C

Radial uniformity: ±0.05°C at -15°C, ±0.07°C at 0°C, ±0.09°C at 110°C

Request a Quote
Download Datasheet

CALsys -15/110 offers medium temperature range from -15°C to 110°C. It is highly stable standard temperature source for calibrating RTD/Thermocouple.

It has peltier elements to generate stable temperature in the well. The comparison volume is a metallic block of aluminium, which has diameter of 24mm and 120mm length.

    CALsys -15/110
Temperature Range at 25°c
-15 to 110°C
Stability ±0.03°C at -15°C
±0.05°C at 0°C
±0.07°C at 110°C
Radial uniformity ±0.05°C at -15°C
±0.07°C at 0°C
±0.09°C at 110°C
Hysteresis 0.02 °C
Immersion depth
120 mm
Insert OD dimensions 24 mm
Resolution 0.1°C
Heating time 10 Min
Cooling time 25 Min ( 110 °C to -15°C)
Resolution 0.1 °C
Display LCD, °C, or °F user-selectable
Method of control Digital self tuned PID Controller
Computer Interface USB
Power Requirement 230 VAC, 500 W(50 Hz)
Size (H x W x D) 425(H) x 230(W) x 305(D) mm
Weight 12 Kg
Calibration Accredited calibration certificate provided (Optional)
Environmental operating conditions 5 °C to 25 °C, 0 % to 90 % RH (non-condensing)
Specifications valid in environmental conditions 5°C … 25°C
Input (Autocal -15/110) Four channels (one master and three test sensors).high-quality universal LEMO connector suitable both for T/C (J, K, N, T, R, S, B type) and Rtd
Software (Autocal -15/110) The calibrator will be provided with software for data recording (Manual Mode ) and Excel genration in AUTO MODE
Data logging (Autocal -15/110)
Data logging facility with logged data export to computer through LAN port ( optional USB)
  • Wide Temperature Range
  • Lightweight, portable
  • Speed
  • Accuracy and performance
  • Accredited calibration
  • Computer Interface
  • Insert
  • Reference Standard RTD
  • Calsoft Software
  • Operational Manual
  • Carry Case
  • NABL accredited calibration certificate (Optional)
  • CE Certified

CONTACT

Let’s Talk About Your Thermal and Cable Needs

Have questions about temperature sensors, heating solutions, or custom cables? We're here to provide expert guidance and tailored solutions. Reach out today, and let’s collaborate on your next project!