CALsys 1200L – High Stability Dry Block Furnace
Part Code: CALsys 1200L
Temperature Range: 300°C to 1200°C
Temperature Resolution: 1.0°C
Stability: ±0.1°C at 300°C, ±0.2°C at 800° C, ±0.35°C at 1200°C
Radial uniformity: ±0.15°C at 300°C, ±0.25°C at 800°C, ±0.4°C at 1200°C
CALsys 1200L offers easy to use temperature calibrator with wide temperature range from 300°C to 1200°C.
It is a highly stable standard furnace for calibrating thermocouples or RTD. This calibrator can be used on site for high temperature calibration and also finds its application in glass, electric power, automotive and material process industry.
The comparison volume is a metallic block of special material, which has a diameter of 37mm and 160mm depth. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment.
Temperature Range |
300 °C to 1200 °C |
Stability |
±0.1°C at 300°C ±0.2°C at 800° C ±0.35°C at 1200°C |
Radial uniformity | ±0.15°C at 300°C ±0.25°C at 800°C ±0.4°C at 1200°C |
Controlling Sensor |
R type duplex |
Stabilization time | 15 to 20 mins |
Immersion depth | 160mm |
Insert OD dimensions | 37 mm |
Method of Control |
Self tuned PID Controller |
Heating time | 1.5 Hrs |
Resolution | 1 °C |
Display | LCD, °C, or °F user-selectable |
Size (H x W x D) | 590(H) x 450(W) x 530(D) mm |
Weight | 55Kg |
Power Requirement | 230 VAC or 110 VAC 50/60 Hz 2.5kW |
PC Interface | LAN/ USB |
Calibration | Accredited calibration certificate provided (Optional) |
Environmental operating conditions | 0 °C to 40 °C |
Input (Autocal 1200L ) | Four channels (one master and three test sensors).high quality universal LEMO connector suitable both for T/C (J, K, N, T, R, S, B type) and Rtd |
Software (Autocal 1200L ) | The calibrator will be provided with software for data recording (Manual Mode ) and Excel genration in Auto Mode |
- Wide Temperature Range
- Simple to use (300°C to 1200°C)
- Accuracy and performance
- Accredited calibration
- Computer Interface
- Insert
- Reference Standard Thermocouple (’N’ Type T/C) (Optional)
- Calsoft Software
- Operational Manual
- Carry Case
- NABL accredited calibration certificate (Optional)
- CE Certified