Enquire Now
Calsys 100 40 1

CALsys -100/40 Portable ,FPSC system based low temperature calibrator

Part Code: CALsys -100/40

Temperature Range: -100 °C to 40 °C

Stability (30min): ±0.04 °C or Better

Radial Uniformity: ±0.08°C or Better

Request a Quote
Download Datasheet

Calsys -100/40 offers easy to use portable low temperature calibrator with temperature range from -100 to 40°C. It is a highly stable standard furnace for calibrating RTD. This calibrator can be used on site for ultra low temperature calibration and also find application in aerospace, oil gas petrochemical, pharmaceutical industry, electric power, automotive and material process industry. The comparison volume is a metallic fixed block of special material, which has a fixed insert with 150mm long.

Low temperature dry block furnace based on FPSC cooling system.This model provides special design isothermal enclosure which can calibrate sensor against the calibrator. Temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment.

SPECIFICATIONS

Temperature Range* -100°C to 40°C
Stability (30 Min) ±0.04 °C or Better
Radial Uniformity ±0.05°C or better
Immersion Depth** 150 mm
Method of Control Self tunned PID controller
Cooling Time 70 Min ( Ambient 25°C to -100°C)
Stabilization Time 15-20 Min
Resolution 0.1 °C (0.01 Optional)
Display LCD, °C, or °F user-selectable
Size (H x W x D) 545(H) x 245(W) x 350(D) mm
Weight 16Kg
Power Requirements 230 VAC or 110 VAC 50/60 Hz 350W
Computer Interface USB/ LAN
Calibration Accredited calibration certificate provided (Optional)
Environmental Operating Conditions 12 °C to 35 °C, 0 % to 90 % RH (non-condensing)
Specifications valid in environmental
conditions
15°C … 30°C
Input (Autocal -100/40) Four channels (one master and three test sensors).high quality universal LEMO connector suitable both for T/C (J,
K, N,T,R,S, B type) and Rtd
Software (Autocal -100/40) The calibrator will be provided with software for data recording (Manual Mode ) and Excel generation in Auto Mode.

* At 25°C Ambient Temperature
**120mm Deep Plus 30 mm top insulation

Note : Customized options available for resolution and insert construction

  • Wide TemperatureRange (-100 to 40°C)
  • Lightweight, portable
  • Speed
  • Accuracy and performance
  • Cooling Technology
  • Accredited calibration
  • Computer Interface
  • Insert
  • Reference Standard RTD
  • Calsoft Software
  • Operational Manual
  • Carry Case