Autocal -30/110 SP
Part Code: Autocal-30/110 SP
- Temperature Range: -25°C to 150°C
- Stability: ±0.01°C or Better
- Resolution: 0.01°C or better
Autocal -30/110 SP easy to use portable low-temperature calibrator with a temperature range from -25 to 150°C. It is a highly stable standard furnace for calibration and also nd application in aerospace, oil gas petrochemical, pharmaceutical industry, electric power, automotive and material process industry.
SPECIFICATIONS
| ITEM DESCRIPTION | SPECIFICATION |
| Function | Metrological purpose (for calibration of Pt 100 and temperature gauges) |
| Temperature Range | -25°C to 150°C |
| Operating temperature | 23°C |
| Accuracy (when internal reference sensor) | ±0.25°C or better |
| Stability | ±0.01°C or Better |
| Stability (with External Reference Sensor) | ±0.01°C or Better |
| Resolution | 0.01°C or better |
| Heating Time | Maximum 50 minutes from ambient to 150°C |
| Cooling Time | Maximum 30 minutes from ambient to -25°C |
| Readout for Master/external reference sensor | It has an inbuilt readout for an externally connected Master/reference RTD sensor (PT100) |
| Readout for Instrument under test (IUT) | It Has an inbuilt readout for instruments under test. |
| Resistance (RTD) | 0-300 ohms (with accuracy 0.009 ohm or better) |
| Inserts Specification | Provision of 4 insert combination of 25mm dia for external reference RTD and DUT (RTD) with options to insert 1(12mm, 3mm, 4mm, 5mm) Insert 2(10mm, 4mm, 8mm, 6mm), Insert 3 (3mm, 4mm, 5mm, 6mm) Insert 4 (16mm, 3mm) |
| Depth (Has sufficient depth to control stem conduction) | Minimum 140 mm |
| Weight | 12 kg |
| Size | 380(H) x 170(W) x 188(D) mm |
| Computer Interface | Ethernet |
| Display | LCD, Unit: °C |
| Inputs on Instrument | Instrument to have connections for RTD (reference probe) 4 wire RTD (IUT) and Thermocouple (K type – IUT). |
| Input Measurement Channels and quantity | (i) RTD (PT100) reference sensor – 1 No. min. (ii) 4 wire RTD – 1 No. min. (iii) K type Thermocouple – 1 No. min. If the input at sl. (ii) and (iii) use a connector plug then 5 nos. of extra connecters for each input to be provided. |
- Wide Temperature Range
- Lightweight, portable
- Accuracy and performance
- Accredited calibration
- Highly accurate & Automatic Dry Block Calibrator for Industrial/Laboratory field use