Autocal 1700L – Automatic High Temperature Dry Block Furnace
Part Code: CALsys 1700L Autocal
Temperature range: 500 °C to 1700 °C
Controlling sensor: B type duplex
Method of Control: Self tunned PID controller
Calibration: Accredited calibration certificate provided
CALsys 1700L Autocal offers easy to use temperature calibrator with a high-temperature range from 500 to 1700°C. It is a highly stable standard furnace for calibrating thermocouples / RTD. This calibrator can be used on-site for high-temperature calibration and also find application in the glass, electric power, automotive and material process industries.
Temperature Range |
500 °C to 1700 °C |
Stability |
±0.5°C at 500°C ±1.0°C at 1000°C ±1.5°C at 1700°C |
Radial Uniformity |
±0.6°C at 500°C ±1.4°C at 1000°C ±1.9°C at 1700°C |
Controlling Sensor |
B type duplex |
Stabilization time | 15 to 20mins |
Immersion depth | 185mm |
Insert OD dimensions | 37 mm |
Method of Control | Self Tuned PID controller |
Heating time | 3 Hrs |
Resolution | 1 °C |
Display | LCD, °C, or °F user-selectable |
Size (H x W x D) | 640(H) x 500(W) x 550(D) mm |
Weight | 130Kg |
Power Requirement | 230 VAC 50/60Hz |
Computer Interface | RS – 232/USB |
Calibration | Accredited calibration certificate provided (Optional) |
Environmental operating conditions | 0 °C to 40 °C, 0 % to 90 % RH (non-condensing) |
Specifications valid in environmental conditions | 13 °C … 33 °C |
Input (CALsys 1700L Autocal) | Four channels (one master and three test sensors).high-quality universal LEMO connector suitable both for T/C (J, K, N, T, R, S, B type) and Rtd |
Software (CALsys 1700L Autocal) | The calibrator will be provided with software for data recording (Manual Mode ) and Excel genration in Auto Mode |
Data logging (CALsys 1700L Autocal) | Data logging facility with logged data export to the computer through LAN port ( optional USB) |
- Wide Operating Range (500 °C to 1700 °C)
- Simple to use
- Accuracy and performance
- Accredited calibration
- Computer Interface
- Insert
- Reference Standard Thermocouple (PT-RT/PT “B type T/C”)
Part No. TTCB – 300 (Optional) - Calsoft Software
- Operational Manual
- Carry Case
- NABL accredited calibration certificate (Optional