FASTCAL 3000 SP High Temperature Black Body Source

Wavelength: 1000 nm or 1600 nm (Pyrometer)

Cavity Temp. Accuracy: ± 0.25 % of set temperature ±1°C

Temperature Stability: ±1°C

Emissivity: 0.995 ±0.005

In stock
SKU
FASTCAL 3000 SP

Part Code: FASTCAL 3000 SP

$0.00

FastCAL 3000 SP is a high temperature blackbody source with a capacity to produce high temperature using high emissivity emitter(graphite tube) and at the same time stabilize at the required temperature within a few minutes.

It is mainly used to calibrate non-contact temperature sensors. Step change in temperature is achievable within seconds. The graphite tube has an emitter diameter of 25mm (1.00") and an emissivity of +0.995 ±0.005. Target temperature is sensed by a non contact temperature sensor/Pyrometer which gives a feedback to PID controller to regulate the cavity temperature precisely to the desired set points. Remote set point is achieved via RS-232 C communication port.

All safety features to protect equipment and operator are easily accessible.

Cavity Closed end graphite tube Length -292mm Heated Length - 127mm
Heating Aperture/Cavity Diameter
Graphite tube with argon gas purge diameter 25mm
Cooling
Water cooling system through chiller unit with all necessary accessories & Tubing nozzle
Purge Gas Dry Argon gas flow with 7-10 LPM
Heating Time
Approx 10 Minute from 600 to 3000°C
Safety Interlocks
Cooling water over temperature, Low purge gas flow, low cooling water flow rate, over current and over temperature protection.
Controlling Sensor
Pyrometer
Method of Control PID controller Eurotherm make(Option - HMI & Data Logging)
Remote Controller Set Point control and temperature monitoring by RS-232/Rs-485
Ambient Temperature 0°C to 50°C
Power 240 VAC, 50Hz
Dimension 1700mm(H) x 900mm(W) x 1200mm(D)
Weight 300 Kg
Cavity Temperature Stability ±.1°C
Cavity Size Varying Adjustment Aperture Size (as per dia needed)
Water DM Water
Controller Unit Display Unit
Calibration certificate NABL/ NPL India/ NMI
  • Water flow failure
  • Increase in water temperature
  • Argon gas flow failure
  • Over current

CE Certified

Write Your Own Review