CALSYS C-4004 PLUS

  • High accuracy of RTD measurement 0.01°C.
  • High accuracy of T/C measurement 0.1°C.
  • 12 Thermocouples B, C, D, E, J, K, N, R, S, T .
  • 12 RTD's PT-100, PT-10, PT-50, PT-500, PT-200, PT-1000, SPRT .
  • Thermocouple reference junction external, internal, and off .
In stock
SKU
CALSYS C-4004 PLUS

Part Code: CALSYS C-4004 PLUS

$0.00

Details

CALSYS C-4004 PLUS is fully characterized for all major sensors (thermocouples such as B, C, D, E, J, K, N, R, S, T and RTD's like PT-100, PT-50, PT-10, PT-200, PT-500, PT-1000, SPRT). The instrument can be used in industries where high accuracy temperature measurement is essential.

CALSYS C-4004 PLUS offers three modes of configuration, which are user-configurable. These three modes are Only TC mode: In this mode, the user can connect a maximum of 12 TC inputs at a time.

Only RTD mode: In this mode, the user can connect a maximum of 12 RTD inputs at a time. Mix Mode (TC + RTD): In this mode, the user can connect a maximum of 6 TC inputs and 6 RTD inputs at a time.

Over-size screw terminal connection blocks are standard for easier connection. Autocalibration facility of multiple channels using PC software.

Specifications

Display
192x64 LCD Graphic Panel with backlight
Input 2 Channel for RTD and Thermocouple via lemo 1- S series connector
Reference Junction

Reference Juncon compensaon may be selected for the following modes :

Automac : Internal reference juncon range 0 to +100 °C

External : Via Pt 100 sensor connected to channel A and B range to 0 to +100 °C

Off : Turns the Reference Juncon OFF = 0° C

Working Temperature 0....50°C rel. humidity Less than 90% non condensing condition
Storage Temperature
-20 to +55 °C
Mains Supply
85-264 VAC / 120-370 VDC 47-63 Hz
Data Logging Up to 4000 single channel Values can be stored along with the date and time
Dimensions 224(H)x350(W)x114(D)mm

 

Key Features

  • High stability
  • High accuracy of RTD measurement 0.01°C
  • High accuracy of T/C measurement 0.1°C
  • High Resolution (0.01/0.001)
  • 2 measuring inputs
  • 12 Thermocouples B, C, D, E, J, K, N, R, S, T
  • 12 RTD's PT-100, PT-10, PT-50, PT-500, PT-200, PT-1000, SPRT
  • Thermocouple reference junction external, internal and off
  • Units °C/ °F
  • Data Logging 4000 values
  • Suitable for 2/3/4 wire RTD

Note : CALSYS C-4004 PLUS gives standard two channel input for TC &
two channel input for RTD. If user wants more than two input
channels for that CALSYS C-4004 PLUS gives inbuilt mulplexer module.