CALsys-196/-80 – Portable Temperature Calibrator
Part Code: Calsys-196/-80
Temperature Range: -190 to -80°C
Stability: ±0.1°C at-180°C, ±0.1°C at -130°C, ±0.1°C at -80
Uniformity: ±0.2°C at-180°C, ±0.15°C at -130°C, ±0.17°C at -80
CALsys -196/-80 offers ultra low temperature range from -180 to -80°C. It is a highly stable standard source for calibrating Thermocouples & RTD.
It has been designed for low temperature range calibration and find application in the glass, pharma, electrical power, automotive & material processing industries.
The comparison volume is a metallic block of special material having dimension 25mm diameter with 300 mm long which is fixed into the liquid nitrogen container for calibration of RTD & thermocouple. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment.
| Temp Range | -180 to -80°C |
| Stability | ±0.1°C at-180°C ±0.1°C at -130°C ±0.1°C at -80 |
| Uniformity | ±0.2°C at-180°C ±0.15°C at -130°C ±0.17°C at -80 |
| Temperature Resolution | 1.0°C |
| Time to Reach Max. Temp | 30 Mins |
| Controlling Sensor | RTD-PT100 |
| Method of Control | Digital self tuned PID Controller |
| Insert Construction | Dia 30 x 300 mm long (2 x 6.5 mm & 2 x 8.5 mm holes) |
| Computer Interface | USB |
| Immersion Depth | 280 mm |
| Time to Reach Max Temp | 30 Mins |
| Operating Temperature | 20 to 45°C |
| Power Requirement | 230 VAC or 110 VAC 50/60 Hz 300 W |
| Dimensions of Container | 580(H) x 280(Dia) |
| Nitrogen Container | 10 Ltr. |
| Dimension of control panel | 165(H)x310(W)x350(D) |
| Weight of control panel | 10 Kg |
| Weight of container | 15 Kg |
- Large Immersion Depths
- Wide Operating Range (-190 to -80°C)
- Lightweight, portable
- High Stability
- Computer Interface
- Accredited calibration
- Insert
- Reference Standard RTD
- Calsoft Software
- Operational Manual
- NABL accredited calibration certificate (Optional)