Autocal 650 SP by Tempsens, leading manufacturer & supplier in UAE. Offers quick, stable, and precise thermocouple calibration for industrial accuracy.
Part Code: CALsys 650 SP Autocal
- Temperature Range: 50°C to 650°C
- Stability: 0.05°C at maximum temperature or better
- Resolution: 0.01°C or better
Autocal 650 SP offers easy to use portable temperature calibrator with a medium temperature range from 50 to 650°C. It is a highly stable standard furnace for calibrating thermocouples I RTD.This calibrator can be used on-site in workshops, test and measurement rooms as well as laboratories. The comparison volume is a metallic block of special material, which has a diameter of 32mm and 140mm long. The Autocal 650 SP is an automatic temperature calibration system for the Thermocouple and RTDs.
SPECIFICATIONS
| ITEM DESCRIPTION | SPECIFICATION |
| Function | Metrological purpose (for calibration of Pt 100 and temperature gauges) |
| Temperature Range | 50°C to 650°C |
| Accuracy (when using internal reference sensor) | 0.35°C or better for 50°C to 420°C 0.5°C or better for above 420°C |
| Stability | 0.05°C at maximum temperature or better |
| Display | LCD or LED |
| Readout for Instrument under test (IUT) | It Has an inbuilt readout for instruments under test. |
| Readout for Master/external reference sensor | It Has an inbuilt readout for an externally connected Master/ reference RTD sensor (PT100) |
| Resolution | 0.01°C or better |
| Heating Time | For the above-mentioned temperature range: Maximum 25 minutes |
| Cooling Time | For the above-mentioned temperature range: Maximum 50 minutes |
| Insert Specification | Provision of 3 insert combination of 32mm dia for external reference RTD and DUT (RTD) with options to insert 1 (3mm, 4mm, 5mm, 6mm, 8mm) Insert 2(10mm, 3mm, 4mm, 12mm), Insert 3(16mm, 4mm, 6mm) |
| Depth (Has sufficient depth to control stem conduction) | Minimum 140 mm |
| Weight | 10 kg |
| Size | 325(H) x 185(W) x 265(D) mm |
| Computer Interface | Ethernet |
| Sensor Fault Protection | Sensor burnout and short protection |
| Power Requirments | 230 V AC, 50 Hz, with IEC Class 1 protection. |
| Inputs on Instruction | Instrument to have connections for RTD (reference probe), 4 wire RTD (IUT) and Thermocouple (K type-IUT). |
- Wide Temperature Range
- Lightweight, Portable
- Accuracy and performance
- Accredited calibration
- Portable, Highly Stable & Automatic Temperature Calibrator for Industrial/Laboratory Field Use